Hochdruckreaktor, Autoklav, Magnetrührdurchführung, Hydrierung

avalon

High pressure system avalon

Here we have created a highly agile series of high-pressure reactors, which we present to you in the small autoclave family - the A-line. The A-line (andorra, apart, avalon, avalon plus) offers a hand when it comes to batch systems or multi-versions of four, eight or more autoclaves in series. The small autoclaves can be heated, cooled or controlled independently of each other.

Specifications

  • Nominal volume ml: 4 x 60, 4 x 100
  • Working volume ml: 20-40, 40-60
  • Working pressure bar: 200, 280, 480
  • Working temperature °C: 200, 250, 300
  • Material WNr. 1.4435 at 200 bar - WNr. 2.4602 at 250 bar - WNr. 1.4980 at 280 bar
  • Union nut as lock M48 x1.5 mm, width across flats SW55
  • Double sealed: O-ring and medium seal metal to metal (cutting ring).
  • Reactor cover resp. magnetic stirrer drive with a torque of 20 Ncm and integrated micromotor 24V/DC
  • Output speed with ball bearing 0-1'300 rpm, with plain bearing 0-1'200 rpm.
  • Regulating shut-off valve for gas supply
  • Regulating shut-off valve for pressure relief
  • Regulating shut-off valve with immersion tube for sampling
  • Pressure gauge in chemical version
  • Bursting disc holder equipped with bursting disc
  • Immersion tube for thermocouple with ø 3 mm (without thermocouple)
  • Temperature sensor Pt100 or type K suitable
  • Stirring element: gas supply stirrer ø 18 x 64 mm
  • Fluid or electric heating
  • Control unit ordino CS400 is the perfect complement to small reactor apart.

Variants/special features

  • Reactor system avalon as multiversion for 4 units. Further units of 8 or more operated in parallel on request.
  • Pressure stages: 200, 250 or 480 bar depending on the choice of material.
  • Temperature stages °C: 200, 250, 300 depending on the choice of material.
  • Material for wetted parts WNr.: 1.4435 (AISI 316L), 2.4602 (HC22) and 1.4980 (AISI 660) as standard. Versions in Hastelloy C276, B3 or titanium size 2 on request.
  • The autoclave vessel is double-sealed: once in the upper area by an O-ring made of Viton, EPDM or Kalrez and at the bottom via a metal-to-metal medium seal (cutting ring).
  • Temperature sensor Pt1000, type K or type N in the immersion tube for measuring the medium temperature (not included in delivery).
  • By opening the locking nut SW12 mm, the output shaft can be removed for cleaning the inside or bearing replacement. Either ball bearings or plain bearings are used, depending on the medium or material used.
  • Two different heating systems are offered. You can choose either an electric or an oil-heated system (double jacket). The latter with inlet and outlet M16 x 1 mm.

Multi-versions of four, eight or more reactors in series

In the basic version, a quadruple chassis with independent heating units is one of the main features of the avalon. Four complete autoclave units andorra can thus be operated in parallel and can be removed from your heating/cooling unit. The avalon reactor system confidently combines reactor units with fully integrated magnetic feedthroughs, micromotors, gas connections and safety attributes. The plug-in connections for heating and thermocouple are located on the rear of the autoclave chassis.